관심표준 등록 : 표준업데이트 시 알림을 받을 수 있습니다.
PDF : 직접 파일 다운로드 및 인쇄 (마이페이지 확인)
PRINT : 인쇄본 우편발송, 2~3일 소요(PDF파일 미제공)
분야 | |
---|---|
적용범위 | 1.1 This test method covers the measurement of the magnetically induced torque produced by the static magnetic field in the magnetic resonance environment on medical devices and the comparison of that torque a user-specified acceptance criterion. 1.2 This test method does not address other possible safety issues which may include, but are not limited to, magnetically induced deflection force, tissue heating, device malfunction, imaging artifacts, acoustic noise, interaction among devices, and the functionality of the device and the MR system. 1.3 The torque considered here is the magneto-static torque due to the interaction of the MRI static magnetic field with the magnetization of the implant. The dynamic torque due to interaction of the static field with eddy currents induced in a rotating device is not addressed in this test method. Torque induced by currents in lead wires is not addressed by this standard. 1.4 The methods in this standard are applicable for MR systems with a horizontal magnetic field. Not all of the methods described in this standard are applicable for use in an MR system with a vertical magnetic field. The Suspension Method and the Low Friction Surface Method require gravity to be orthogonal to the magnetically induced torsion and may not be performed using a vertical magnetic field. The Torsional Spring and Pulley Methods can be adapted to work in a vertical magnetic field, however the example apparatus are not appropriate for use in a vertical magnetic field. The Calculation Based on Measured Displacement Force Method is independent of the MR system and thus could be used for an MR system with a vertical magnetic field. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee. |
국제분류(ICS)코드 | |
페이지수 | 12 |
Edition | 17 |
No. | 표준번호 | 표준명 | 발행일 | 상태 |
---|---|---|---|---|
1 | ASTM F2213-17 REDLINE | Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | 2017-09-01 | 표준 |
2 | ASTM F2213-17 | Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | 2017-09-01 | 표준 |
3 | ASTM F2213-06(2011) | Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | 2011-10-01 | 구판 |
4 | ASTM F2213-06 | Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | 2006-05-01 | 구판 |
5 | ASTM F2213-04 | Standard Test Method for Measurement of Magnetically Induced Torque on Passive Implants in the Magnetic Resonance Environment | 2004-01-01 | 구판 |
6 | ASTM F2213-02 | Standard Test Method for Measurement of Magnetically Induced Torque on Passive Implants in the Magnetic Resonance Environment | 2002-11-10 | 구판 |
관련상품이 존재하지 않습니다.
ASTM F2182-19e2 - Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging 상세보기
ASTM F2052-21 REDLINE - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment 상세보기
ASTM F2503-20 REDLINE - Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment 상세보기
ASTM F2052-15 REDLINE - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment 상세보기
ASTM F2119-07(2013) - Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants (Withdrawn 2022) 상세보기
IEC TS 63134:2020 - Active assisted living (AAL) use cases 상세보기
IEC 60034-5:2020 RLV - Rotating electrical machines - Part 5: Degrees of protection provided by the integral design of rotating electrical machines (IP code) - Classification 상세보기
KS B ISO TS 25740-1 - 에스컬레이터 및 무빙워크에 대한 안전요건 — 제1부: 세계공통 필수 안전요건(GESRs) 상세보기
KS B ISO TS 8100-21 - 승객 및 화물 운송용 엘리베이터 —제21부: 세계공통 필수안전요건(GESRs)을 충족하는 세계공통 안전 파라미터(GSPs) 상세보기
KS C IEC TS 62872 - 산업 시설과 스마트 그리드 사이의 산업 공정 측정, 제어 및 자동화 시스템 인터페이스 상세보기